What is Rapicide used for?

RAPICIDE™ High-Level Disinfectant and Sterilant is a fast-acting, non-staining disinfectant that requires no mixing or activation. This reusable glutaraldehyde disinfectant is designed for use on flexible endoscopes in automated endoscope reprocessors. Its fast contact time allows for quick endoscope turnaround time.

Who makes Rapicide?

Cantel
RAPICIDE™ OPA/28 High-Level Disinfectant | Cantel.

What is Rapicide Opa?

RAPICIDE™ OPA/28 is an ortho-Phthalaldehyde based high-level disinfectant that features a long reuse life. With the same contact time and temperature for manual disinfection as in an automated endoscope reprocessor, RAPICIDE OPA/28 provides fast turnaround time for endoscopes.

How do you use Rapicide?

Immerse device completely, filling all lumens or channels if present, with RAPICIDE OPA/28 High-Level Disinfectant solution at or above 0.35% OPA for a minimum of 5 minutes at room temperature (minimum 20°C). Monitor the MRC of the solution with RAPICIDE OPA/28 Test Strips prior to each reprocessing cycle.

Is Rapicide flammable?

Flammability (solid, gas) : Not flammable. Vapour pressure : No data available. Relative vapour density at 20 °C : No data available. Relative density : 1.013 Solubility : No data available.

What is Rapicide made of?

RAPICIDE™ PA High-Level Disinfectant is a single-use, peracetic acid-based solution that is environmentally responsible with proven materials compatibility. It is designed for use in the ADVANTAGE PLUS™ Pass-Thru, ADVANTAGE PLUS™ or DSD EDGE™ Automated Endoscope Reprocessors.

Who manufactures cidex Opa?

Johnson & Johnson company
CIDEX products are manufactured by Advanced Sterilization Products, a division of Ethicon (a Johnson & Johnson company). CIDEX OPA is an effective medical instrument disinfectant and it is not to be used on skin.

Is cidex flammable?

Emergency overview: Flammable. Properties affecting health: Toxic if swallowed. Irritating to eyes, skin and respiratory tract.

How long is Opa good for?

2 years
The shelf life of an un- opened bottle of CIDEX OPA Solution is 2 years. The solu- tion requires NO activation. After opening the bottle, pour CIDEX OPA Solution into a CIDEX Solution Tray or appropriate container. If there is still solution left in the bottle, the bottle can be stored up to 75 days.

How long are cidex OPA test strips good for?

The shelf life (expiration date) for the unopened CIDEX OPA Solution Test Strips is stamped on the immediate container lable. When opening the bottle for the first time, record the date opened in the space provided on the label. PRECAUTIONS: Do not use any remaining strips 90 days after opening the bottle.

Is Opa toxic?

The toxicological properties of CIDEX OPA Solution have been studied extensively. Results demonstrate that the solution is safe when used as directed. A summary of results from the toxicity tests performed appears below.

When to use rapicide OPA / 28 Indicator test strips?

The indicator test strips are used for testing RAPICIDE OPA/28 High-Level Disinfectant, and ortho-phthalaldehyde based disinfectant available for use in an automated endoscope reprocessor (AER) or for manual disinfection.To Order:Contact your local distributor or RAPICIDE™ OPA/28 Test Strips | Cantel

What kind of disinfectant is rapicide OPA 28?

RAPICIDE™ OPA/28 High-Level Disinfectant is a fast-acting, long lasting, highly compatible high-level disinfectant ensuring a safe and healthy environment for patients and staff. RAPICIDE™ OPA/28 High-Level Disinfectant | Cantel

What do you need to know about rapicide test strips?

Periodic testing ensures the test strips are being properly stored and consistently providing accurate MRC readings. A quality assurance protocol is outlined on the instructions for use provided with every box of Rapicide®OPA/28 Test Strips. I am noticing the need to add additional chemistry to my reservoirs after using Rapicide®OPA/28.

How long can rapicide be used for Aer?

SANIProZymeTMEnzymatic Detergent. In AER application, it’s also important to have annual AER Preventative Maintenance and to maintain a consistent chemistry temperature in the reservoir and basin. Rapicide OPA/28 is reusable for up to 28 days provided the MRC is verified to be above 0.35%.